Admittance and conductance in cardiac performance
Purposedetermine the instantaneous volume of the ventricle

There is an established practice of using the electrical conductance of blood ( PV loops) in heart ventricles to determine the instantaneous volume of the ventricle. This technique involves inserting a tetra-polar catheter into the ventricle and measuring conductance. This measured conductance is a combination of blood and muscle and various techniques are used to identify the blood conductance from the total measured conductance. Blood conductance can then be converted to volume using a linear (Baan) or a non-linear (Wei) relationship that relates conductance to volume.

Traditional conductance technology

This approach is based on the idea that the total conductance, G, of a fluid between two electrodes is a function of the fluid's conductivity (reciprocal of resistivity) and volume.[ citation needed]

In cardiology, a tetra-polar catheter is inserted into the ventricle and a constant current (I) is applied across the two outer electrodes. This generates an electrical field within the ventricle and the two inner electrodes measure a voltage generated due to the electric field. This measured voltage (V) is used to determine conductance through a modified version of Ohm's Law. Conductance (G) is the reciprocal of resistance (R) which changes the standard Ohm's equation from V=IR to V=I/G.[ citation needed]

Conductance is then related to blood volume though Baan's equation. [1] When used in cardiology, the electric field generated is not limited to the blood (the fluid of interest) but also penetrates the heart wall, giving rise to additional conductance often called "parallel conductance" or "muscle conductance", Gm which must be removed.[ citation needed]

Various techniques have been attempted to remove the Gm contribution with varying degrees of success. The most common method is the hypertonic saline technique which involves injecting a bolus of hypertonic saline into the ventricle to alter blood conductivity without affecting the surrounding muscle. Another less commonly used technique involves evacuating the ventricle of blood and measuring muscle conductance alone with a conductance catheter. Clearly both techniques are unreliable, somewhat invasive and fail to account for the continuous variation in Gm over the cardiac cycle. [2]

Improved admittance technology

The Admittance technique is an improvement over the Conductance technique for the real-time removal of muscle conductance Gm. Blood and muscle respond to alternating (AC) electrical currents very differently. Blood is purely resistive while muscle has both resistive and capacitive properties. The fixed charges in muscle cells create a significant reactance that causes a phase shift (time delay) in the measured signal, relative to the excitation signal. Admittance technology uses this phase shift to determine the instantaneous muscle conductance and remove it from the total measured conductance.[ citation needed]

Removal of Gm

The total Admittance (Y), of the blood filled ventricle is given by Y = Gb + Gm + iωCm where:[ citation needed]

  • Gb is the measured conductance of the blood (the desired signal)
  • Gm is the measured conductance of the cardiac muscle (unwanted signal)
  • Cm is the measured capacitance of the cardiac muscle
  • ω is the angular frequency of the excitation signal

The signals Gm and Cm are both properties of cardiac muscle and vary in a fixed ratio. Thus the ratio of Gm to Cm is equal to the ratio of muscle conductivity (σ) to muscle permittivity (ε). The ratio σ/ε is the constant of proportionality. Although both σ and ε are functions of the health of the heart tissue, they are relatively constant for short periods of time.[ citation needed]

Using this proportionality, one can rewrite the equation for Gm as Gm = (σ/ε)Cm

Note that the imaginary component of Y depends only on the amount of muscle in the field of the catheter. This makes it easy to isolate by measuring the phase shift, φ, of the measured signal:[ citation needed]
cos(φ) = (Gb + Gm)/Y
sin(φ) = ωCm/Y
Hence, Cm = Y.sin(φ)/ω
also, Gm = (σ/ε)Cm

Thus, blood conductance is determined as Gb = Y.cos(φ) - Gm

Wei's equations can be applied to this calculated blood conductance Gb to obtain blood volume. [3] Unlike Baan's equation, Wei's equation takes into account the non-linear nature of the electrical field and the dynamic nature of the cardiac cycle to give a more accurate representation of the blood volume. [4]

Experimental results

Admittance technique involves the measurement of both phase angle and total conductance in the ventricle. Thus, it is possible to observe how the parallel conductance (muscle conductance) varies throughout the cardiac cycle. A plot showing both the blood and muscle contribution are shown in the figure.[ citation needed]

References

  1. ^ Baan J, Jong TT, Kerkhof PL, et al. (June 1981). "Continuous stroke volume and cardiac output from intra-ventricular dimensions obtained with impedance catheter". Cardiovasc. Res. 15 (6): 328–34. doi: 10.1093/cvr/15.6.328. PMID  7296590.
  2. ^ Krenz M (December 2009). "Conductance, admittance, and hypertonic saline: should we take ventricular volume measurements with a grain of salt?". J. Appl. Physiol. 107 (6): 1683–4. doi: 10.1152/japplphysiol.01089.2009. PMID  19797683.
  3. ^ Wei CL, Valvano JW, Feldman MD, Pearce JA (October 2005). "Nonlinear conductance-volume relationship for murine conductance catheter measurement system". IEEE Trans Biomed Eng. 52 (10): 1654–61. CiteSeerX  10.1.1.124.7182. doi: 10.1109/TBME.2005.856029. PMID  16235651. S2CID  12033677.
  4. ^ Porterfield JE, Kottam AT, Raghavan K, et al. (December 2009). "Dynamic correction for parallel conductance, GP, and gain factor, alpha, in invasive murine left ventricular volume measurements". J. Appl. Physiol. 107 (6): 1693–703. doi: 10.1152/japplphysiol.91322.2008. PMC  2793194. PMID  19696357.

Further reading

  • Wei CL, Valvano JW, Feldman MD, Nahrendorf M, Peshock R, Pearce JA (August 2007). "Volume catheter parallel conductance varies between end-systole and end-diastole". IEEE Trans Biomed Eng. 54 (8): 1480–9. doi: 10.1109/TBME.2007.890732. PMID  17694869. S2CID  15306967.
  • Clark JE, Kottam A, Motterlini R, Marber MS (2009). "Measuring left ventricular function in the normal, infarcted and CORM-3-preconditioned mouse heart using complex admittance-derived pressure volume loops". J Pharmacol Toxicol Methods. 59 (2): 94–9. doi: 10.1016/j.vascn.2008.10.007. PMID  19059354.
  • US 7925335, Feldman, Marc D.; Pearce, John A.; Valvano, Jonathan W.; Wei, Chia-Ling, "Method and Apparatus for Determining Cardiac Performance in a Patient with a Conductance Catheter", published April 12, 2011 
From Wikipedia, the free encyclopedia
Admittance and conductance in cardiac performance
Purposedetermine the instantaneous volume of the ventricle

There is an established practice of using the electrical conductance of blood ( PV loops) in heart ventricles to determine the instantaneous volume of the ventricle. This technique involves inserting a tetra-polar catheter into the ventricle and measuring conductance. This measured conductance is a combination of blood and muscle and various techniques are used to identify the blood conductance from the total measured conductance. Blood conductance can then be converted to volume using a linear (Baan) or a non-linear (Wei) relationship that relates conductance to volume.

Traditional conductance technology

This approach is based on the idea that the total conductance, G, of a fluid between two electrodes is a function of the fluid's conductivity (reciprocal of resistivity) and volume.[ citation needed]

In cardiology, a tetra-polar catheter is inserted into the ventricle and a constant current (I) is applied across the two outer electrodes. This generates an electrical field within the ventricle and the two inner electrodes measure a voltage generated due to the electric field. This measured voltage (V) is used to determine conductance through a modified version of Ohm's Law. Conductance (G) is the reciprocal of resistance (R) which changes the standard Ohm's equation from V=IR to V=I/G.[ citation needed]

Conductance is then related to blood volume though Baan's equation. [1] When used in cardiology, the electric field generated is not limited to the blood (the fluid of interest) but also penetrates the heart wall, giving rise to additional conductance often called "parallel conductance" or "muscle conductance", Gm which must be removed.[ citation needed]

Various techniques have been attempted to remove the Gm contribution with varying degrees of success. The most common method is the hypertonic saline technique which involves injecting a bolus of hypertonic saline into the ventricle to alter blood conductivity without affecting the surrounding muscle. Another less commonly used technique involves evacuating the ventricle of blood and measuring muscle conductance alone with a conductance catheter. Clearly both techniques are unreliable, somewhat invasive and fail to account for the continuous variation in Gm over the cardiac cycle. [2]

Improved admittance technology

The Admittance technique is an improvement over the Conductance technique for the real-time removal of muscle conductance Gm. Blood and muscle respond to alternating (AC) electrical currents very differently. Blood is purely resistive while muscle has both resistive and capacitive properties. The fixed charges in muscle cells create a significant reactance that causes a phase shift (time delay) in the measured signal, relative to the excitation signal. Admittance technology uses this phase shift to determine the instantaneous muscle conductance and remove it from the total measured conductance.[ citation needed]

Removal of Gm

The total Admittance (Y), of the blood filled ventricle is given by Y = Gb + Gm + iωCm where:[ citation needed]

  • Gb is the measured conductance of the blood (the desired signal)
  • Gm is the measured conductance of the cardiac muscle (unwanted signal)
  • Cm is the measured capacitance of the cardiac muscle
  • ω is the angular frequency of the excitation signal

The signals Gm and Cm are both properties of cardiac muscle and vary in a fixed ratio. Thus the ratio of Gm to Cm is equal to the ratio of muscle conductivity (σ) to muscle permittivity (ε). The ratio σ/ε is the constant of proportionality. Although both σ and ε are functions of the health of the heart tissue, they are relatively constant for short periods of time.[ citation needed]

Using this proportionality, one can rewrite the equation for Gm as Gm = (σ/ε)Cm

Note that the imaginary component of Y depends only on the amount of muscle in the field of the catheter. This makes it easy to isolate by measuring the phase shift, φ, of the measured signal:[ citation needed]
cos(φ) = (Gb + Gm)/Y
sin(φ) = ωCm/Y
Hence, Cm = Y.sin(φ)/ω
also, Gm = (σ/ε)Cm

Thus, blood conductance is determined as Gb = Y.cos(φ) - Gm

Wei's equations can be applied to this calculated blood conductance Gb to obtain blood volume. [3] Unlike Baan's equation, Wei's equation takes into account the non-linear nature of the electrical field and the dynamic nature of the cardiac cycle to give a more accurate representation of the blood volume. [4]

Experimental results

Admittance technique involves the measurement of both phase angle and total conductance in the ventricle. Thus, it is possible to observe how the parallel conductance (muscle conductance) varies throughout the cardiac cycle. A plot showing both the blood and muscle contribution are shown in the figure.[ citation needed]

References

  1. ^ Baan J, Jong TT, Kerkhof PL, et al. (June 1981). "Continuous stroke volume and cardiac output from intra-ventricular dimensions obtained with impedance catheter". Cardiovasc. Res. 15 (6): 328–34. doi: 10.1093/cvr/15.6.328. PMID  7296590.
  2. ^ Krenz M (December 2009). "Conductance, admittance, and hypertonic saline: should we take ventricular volume measurements with a grain of salt?". J. Appl. Physiol. 107 (6): 1683–4. doi: 10.1152/japplphysiol.01089.2009. PMID  19797683.
  3. ^ Wei CL, Valvano JW, Feldman MD, Pearce JA (October 2005). "Nonlinear conductance-volume relationship for murine conductance catheter measurement system". IEEE Trans Biomed Eng. 52 (10): 1654–61. CiteSeerX  10.1.1.124.7182. doi: 10.1109/TBME.2005.856029. PMID  16235651. S2CID  12033677.
  4. ^ Porterfield JE, Kottam AT, Raghavan K, et al. (December 2009). "Dynamic correction for parallel conductance, GP, and gain factor, alpha, in invasive murine left ventricular volume measurements". J. Appl. Physiol. 107 (6): 1693–703. doi: 10.1152/japplphysiol.91322.2008. PMC  2793194. PMID  19696357.

Further reading

  • Wei CL, Valvano JW, Feldman MD, Nahrendorf M, Peshock R, Pearce JA (August 2007). "Volume catheter parallel conductance varies between end-systole and end-diastole". IEEE Trans Biomed Eng. 54 (8): 1480–9. doi: 10.1109/TBME.2007.890732. PMID  17694869. S2CID  15306967.
  • Clark JE, Kottam A, Motterlini R, Marber MS (2009). "Measuring left ventricular function in the normal, infarcted and CORM-3-preconditioned mouse heart using complex admittance-derived pressure volume loops". J Pharmacol Toxicol Methods. 59 (2): 94–9. doi: 10.1016/j.vascn.2008.10.007. PMID  19059354.
  • US 7925335, Feldman, Marc D.; Pearce, John A.; Valvano, Jonathan W.; Wei, Chia-Ling, "Method and Apparatus for Determining Cardiac Performance in a Patient with a Conductance Catheter", published April 12, 2011 